/ Home / Chopping Systems / CH-SYS-1C
PLD-1S

CH-SYS-1C

Resonant Chopper Locked To External Clock

Features and Advantages

  • Phase locked to an external clock input
  • 10 Hz to 16 KHz frequency range
  • Both sine wave and square wave reference signals
  • Clock stability +/-50 PPM
  • 0.01% phase stability
  • Position output
  • 0.01% or better amplitude stability

Product Options

PLD-1S-PC
Board level driver that requires an external +/-12V to +/-15V dc power supply.
PLD-1S-110
Cased driver (12" × 10" × 3.8") operating from a line voltage of 110V AC.
PLD-1S-220
Cased driver (12" × 10" × 3.8") operating from a line voltage of 220V AC.
PLD-1S-110/220
Cased driver (12" × 10" × 3.8") with a line switch selector for operating from a line voltage of 110V or 220V AC.
  • Description
  • Drawings
  • Documents

The PLD-1C driver phase locks any tuning fork chopper (type CH, except for the subminiature CH-25) to an external clock input. The system can be used as a "stand alone" unit or can be incorporated into a larger instrument or a system. Although thetuning fork chopper can be used in a large temperature range this driver is not recommended for use in temperature sensitive applications. To lock the chopper to an external clock signal you need to specify the exact clock frequency (to 4 decimal places). The phase relationship to the clock is factory set to customer's requirements (0° to 360°).

CLOCK REQUIREMENTS

EXTERNAL CLOCK STABILITY+/-50 PPM
EXTERNAL CLOCK ACCURACY100 PPM

FRONT PANEL CONTROLS

POWERPower switch to turn the drive "ON"
CLOCK IN(PUT)External clock input (TTL or sine wave), BNC connector
LOCKED MODEChopper vanes are phase locked to the clock signal
OSCILLATOR MODEThe chopper is self oscillating at its resonant frequency (not locked to the clock input
PHASE CONTROL POTPhase adjustment of the chopper in relationship to the clock, +/-45° min.
POSITION MONITORVanes position output, BNC connector
AMPLITUDE CONTROLChopping amplitude adjustment POT
OUTPUTOutput connector to interconnect to chopper

SPECIFICATIONS

Frequency range5 Hz to 16 KHz
External clock signalTTL level, sine or square wave (1V PTP to 20V PTP)
External clock stability+/-50 PPM
Chopper amplitude stability 0.01% or better (not locked to the clock input)
Position monitorAnalog position output, +/-5V max. 1kOhm max load
Phase adjustment range+/-45° min.
Phase stability0.01%
Phase relationshipFactory set to customer specifications
Operating temperature rangeRoom temperature only
Power input110V ac or 220V ac, 50-60 Hz, 20W

ORDERING INFORMATION
A) CHOPPER INFORMATION:
TYPE [CH-10 or CH-20]; DUTY CYCLE[%]; VANE [B=bright or D=dark]; FREQUENCY [Hz]
Example: PART NO. CH10-90D-825. This part number specifies the model CH-10 chopper, with 90% duty cycle, dark vanes and a 825 Hz operating frequency.

B) DRIVER INFORMATION: